Product model:

Brief description:The three-dimensional beam analysis system of the radiotherapy water tank is a modern three-dimensional water tank system. Relevant measurements for accelerator QA. The system's correct position indicator and perfectly level mechanism significantly reduces setup time. Aluminum and stainless steel components ensure refinement and robustness. Microprocessor controlled, the unit features a dual-channel electrometer and control software with reversible polarity and automatic distance adjustment for quick and easy system setup.

  • Manufacturer strength

    Manufacturer Strength
  • Valid warranty

    Valid Warranty
  • Quality assurance

    Quality Assurance

Products

PRODUCT CATEGORY

Detailed introduction

Overview




The three-dimensional beam analysis system of the radiotherapy water tank is a modern three-dimensional water tank system. for acceleratorRelevant measures for QA. The system's correct position indicator and perfectly level mechanism significantly reduces setup time. Aluminum and stainless steel components ensure refinement and robustness. Microprocessor controlled, the unit features a dual-channel electrometer and control software with reversible polarity and automatic distance adjustment for quick and easy system setup. The combination of continuous scanning mode and step scanning mode completes the analysis of measurement data in a shorter measurement time.

1. Overview of the device for automatically scanning water tanks in three dimensions

The radiotherapy automatic scanning water tank system adopts automated measurement and control technology, which can quickly, conveniently and accurately scan and measure various types of medical linear accelerators.The radiation field dose distribution parameters of X-rays and electron beams can also be measured for the beam parameters of the MM50 cyclotron, X-knife and 60Co remote treatment machine.

In order to adapt to the requirements of more institutions and customers.

2. Precision mechanical structure

By using a precision mechanical transmission system and a multi-subdivision stepper motor, the system's scanning positioning accuracy is better than0.1mm。

Automatic scanning system: multi-segment stepper motor stainless steel guide rail and slider.

3. Analysis system software




of the systemThe "ray beam analysis system software package" is fully functional and can collect and analyze data, automatically display measured graphics, and calculate various important parameters. It can also convert data into format files required by the radiotherapy planning system. It can adjust the measurement speed, can continuously measure 10 layers of Profile, can scan PPD off center, and can smooth the measurement curve.

Main technical parameters

water tank

External dimensions of water tank (LxWxH):

680mm x407mm x 350 mm

Scan range (LxWxH):

520 mm x140 mmx200 mm

Location resolution:

0.1 mm

Position accuracy:

±0.1 mm

Position repeatability:

± 0.1 mm

Scan speed:

50 mm/s

Approximate volume:

200 I

wall thickness/Materials:

15 mm /acrylic

Weight:

45 kq

Universal control unit (CCU)

Maximum resolution: Measuring range:

0.5 fAat0.4 nA full scale; 5supplyat 40 nA full scale; 0.5 pA at 4 uA full scale 0.4 nA/40 nA/4uA

Leakage current:

typically<1 fA

Time constant:

20 ms

Bias voltage range:

± 50 through ± 500 V

Trigger interface:

RS 485 (custom specific)

Common interface:

ETHERNET (100BaseT)

Main power supply:

100-240VAC110%: 50/60 Hz

LDA-99SC

Number of semiconductor detectors:

99

Semiconductor detector spacing:

5 mm(center to center)

Semiconductor detector specifications

Silicon type:

Hi-pSitype diode detector

Piece size:

2.45 mm x 2.45 mm

Effective volume thickness:

2.0 mm

Sensitivity:

35 gy/nC

Valid measuring points:

<1 mm

Positioning in the water tank:

0, 45, 90and135degree

emXX

Number of channels:

99+1

Measurement mode:

Dose, dose rate (including real-time), simultaneous measurement of all channels

(Dose/dose rate) double pass

Input range:

0nA to500nA

Accuracy:

± 1 %for currents > 10 pA

Resolution:

100 fC(The charge value reaches0.2 mCtime)

Analog to digital conversion:

120strip synchronized analog-to-digital converter

Size(LxWxH):

370mm x285mm x 130 mm

The best computer configuration

Operating system:

Windows XPProfessional version,Service Pack 2, 32 bit, Windows Vista™.final version,32 bit

Processor:

Dual-core processor (or equivalent),2GHzor higher

The above technical parameters are subject to change without further notice.。