KARRAY / PRODUCT DETAILS

DOSIMAX single-function X-ray machine dose quality control detector

Survive in development, continuously optimize, and promote rapid development of enterprises with good reputation and scientific management

DOSIMAX single-function X-ray machine dose quality control detector image

DOSIMAX single-function X-ray machine dose quality control detector

Product model:

Brief description:DOSIMAX single-function X-ray machine dose quality control detector complies with IEC 61674 standard and is used for stability testing of X-ray film machines and fluoroscopy systems. Different types of semiconductor detectors (RQA or RQM) can be used. RQA detectors are used for quality control testing of radiography, dental photography, CR, DR devices. Voltage: 50-150KV Dose: 200nGy-9999mGy Dose rate: 80nGy/s-70mGy/s Exposure time: 1ms-9999s

  • Manufacturer strength

    Manufacturer Strength
  • Valid warranty

    Valid Warranty
  • Quality assurance

    Quality Assurance

Products

PRODUCT CATEGORY

Detailed introduction

germanyIBA  DOSIMAX single-function X-ray machine dose quality control detector

DOSIMAX single-function X-ray machine dose quality control detectorComplies with IEC 61674 standard and is used for stability testing of X-ray cameras and fluoroscopy systems.

Different types of semiconductor detectors are available (RQA or RQM)

RQA detector

used forQuality control testing of radiography, dental photography, CR, and DR devices

Voltage:50-150KV

Dosage:200nGy-9999mGy

Dose rate:80nGy/s-70mGy/s

Exposure time:1ms-9999s

RQM detector

single functionX-ray machine dose quality control detector is used for quality control testing of mammography X-ray machines

Voltage:25-35KV

Dosage:500nGy-9999mGy

Dose rate:1.5μGy/s-300mGy/s

Exposure time:1ms-9999s

CT ionization chamber

Measurement indicators:

dose, dose rate, time,CT dose index, CT dose length product

Different types of semiconductor detectors are available (RQA or RQM)

RQA detector

used forQuality control testing of radiography, dental photography, CR, and DR devices

Voltage:50-150KV

Dosage:200nGy-9999mGy

Dose rate:80nGy/s-70mGy/s

Exposure time:1ms-9999s

RQM detector

for breastQuality control inspection of X-ray machines

Voltage:25-35KV

Dosage:500nGy-9999mGy

Dose rate:1.5μGy/s-300mGy/s

Exposure time:1ms-9999s

CT ionization chamber

Measurement indicators:

dose, dose rate, time,CT dose index, CT dose length product

DCT10-RS Pencil Ionization Chamber CT Probe: Analysis software can automatically report CTDI values

Total effective length:100mm

Effective measurement volume: 122.50000000000001px3

Dosage:10nGy – 9999mGy

Dose rate:100nGy/s – 120mGy/s

Exposure time:1 ms – 199999 s

leakage ray: < +/-4x10 -15A

Measuring range: 100 kV – 150 kV

scale factor (typ.): ND,K=70mGy*cm/nC(120kV/HWD4.05 mm Al)。

Product consultation

Message box

Please enter the calculation result, such as: three plus four = 7

After submission, we will receive your requirements. You can also download the requirement file or call 021-20989890 directly.