German BERTHOLD LB6411 neutron probe

Survive in development, continuously optimize, and promote rapid development of enterprises with good reputation and scientific management

German BERTHOLD LB6411 neutron probe image

German BERTHOLD LB6411 neutron probe

Product model:

Brief description:It has the highest sensitivity and is very suitable for measuring in very weak neutron radiation fields. ICRP Publication No. 60 recommended the use of a new radiation unit. Bertholdt Company and Karlsruhe Company jointly developed the neutron dose rate meter LB6411. The system probe uses the patented He-3 proportional counter tube.

  • Manufacturer strength

    Manufacturer Strength
  • Valid warranty

    Valid Warranty
  • Quality assurance

    Quality Assurance

Products

PRODUCT CATEGORY

Related articles

RELATED ARTICLES

Detailed introduction

germanyBERTHOLD LB6411 Neutron Probe

Product introduction:

l It has extremely high sensitivity and is very suitable for measurement in very weak neutron radiation fields.

l ICRP Publication No. 60 recommended the use of new radiation dose units. Bertholdt and Karlsruhe jointly developed the neutron dose rate meter LB6411

l The system probe adopts patentedHe-3 proportional counter tube

l The counting tube is placed in diameter 250mm moderating ball center

l The geometric position of the counting tube is based on the Monte Carlo simulation program (MCNP) calculation results and obtained a patent

l The instrument has a very high sensitivity, and everynSv is about 3 counts, which is very suitable for measurements in very weak neutron radiation fields.

l Can be used as a portable dosimeter or as a fixed dose monitor

l The monoenergetic neutron fluence response experiment was conducted at the Braunschweig Institute of Physics in Germany.

Features:

l This instrument has everyNSV has high sensitivity of about 3 counts

l Can be used as a portable instrument or as a fixed monitor

l Calibration of neutron surrounding dose equivalentsH *(10)

l The thermal energy of neutron detection reaches20 MeV

l energy dependence±30% between 50 keV and 10 MeV

l Extremely high sensitivity and therefore low measurement times in weak neutron fields

 

Technical parameters:

System host

LB1230

Display

High contrast LCD dot matrix display, resolution32 x 84 pixels; with backlight, 4-digit display of measurement results, automatic range expansion

Operation

1 switch key, 4 operation keys

Probe connection

8-pin socket Fischer 04, spiral cable LB 75576 (standard)

temperature range

-15℃ - +50℃

Size

145 mm (H) x 170 mm (W) x 45 mm (D)

weight

approx.0. 8kg (including battery)

power supply

3 x IEC-R14 or 3 x rechargeable Varta NiCd #5014 batteries

battery life

UseR14 >150 hours

LB6411 Neutron Detector

Measured quantity

Neutron surrounding dose equivalentH*(10)

Measuring range

30 nSv/h ~ 100 mSv/h

Neutron energy range

Thermal neutrons ~ 20 MeV

scale source

Cf-252 naked neutron source

fluence response

1.09 cm2

scale factor

1.27 μSv/h / cps

response function

The fluence and energy response functions can be found in the manual.

Energy dependence

±30% between 50 keV ~ 10 MeV

gamma sensitivity

reading less than40 μSv/h(10mSv/h Cs-137)

temperature range

-10℃~ 50℃

Moderating ball diameter

625px