Product model:

Brief description:The Compton Suppression System reduces the Compton platform height for low background counting systems, thereby increasing minimum detectable activity and overall spectral quality - particularly useful for small volume samples such as air filters and petri dishes

  • Manufacturer strength

    Manufacturer Strength
  • Valid warranty

    Valid Warranty
  • Quality assurance

    Quality Assurance

Products

PRODUCT CATEGORY

Detailed introduction

The Compton Suppression System reduces the Compton platform height for low background counting systems, thereby increasing the minimum detectable activity and overall spectral quality - particularly useful for small volume samples such as air filters and Petri dishes.

ORTEC offers dozens of Compton suppression systems within the range to meet the demanding requirements of extremely low background counting applications.

ORTEC evaluated various HPGe detectors on a range of performance parameters such as efficiency, resolution, and peak-to-scan ratio to determine the best combination of instruments to achieve the best overall performance.
  • information +

    • Download link Compton Suppression Counting System Brochure
    • Download link Compton Suppression Counting System Brochure (A4)
  • More information +


    Compton suppression system basics
    In a typical low background application system, great efforts are made to reduce the inherent radioactivity of the counting system components (detector, lead shield, and air in the shield chamber). In high-purity germanium spectroscopy systems, these low-background measurement systems tend to have less cosmic ray background and less natural environment background.

    Compton suppression systems are designed to reduce plateau counts in these typical counting systems. While low-background systems remove most of the sources of increased background, they do not resolve the contribution of the continuum below the peak: Compton scattering events. Compton scattering occurs when the entire energy of an incident photon is not absorbed by the HPGe detector, while the portion of the energy leaving the detector cannot be calculated. In the gamma ray spectrum, this partial count appears below the peak of the Compton continuum plateau and is a random event.

    The ratio of the peak height to the average height of the Compton continuum is called the peak-to-Compton ratio (P/C). In standard HPGe detectors, the peak-to-combustion ratio of Co60 at 1.33 MeV is usually between 40:1 and 60:1, and the peak-to-combustion ratio of larger detectors can be close to 100:1.

    Because the escaping energy is emitted as photons, it is possible to collect this ray with another detector. This is usually achieved using a large crystal made of a relatively cheap material (such as NaI), which becomes a coincidence detector. By correlating the count events in the HPGe and coincidence detectors through time electronics, the count time in the coincidence detector can be used to discriminate between simultaneous count events in the HPGe detector and discard them. The result was the suppression of Compton's platform. In a Compton suppression system, a peak-to-scan ratio of over 1300:1 can be obtained using an N-type detector with 60% relative efficiency. This will result in a reduction of about 10 times in platform height and more than 3 times in MDA value.

    Selection Factors for HPGe Detectors in Compton Suppression Systems
    The effectiveness of the Compton suppression system depends on the ability to collect photons scattered by the HPGe detector. Because a photon has the potential to interact with every material it encounters, as little material as possible must be used between the effective volume of the HPGe crystal and the detector crystal. Relevant materials include:

      Features that should be selected  Features that should not be selected
    HPGe external contact grade Ultra-thin external contact grade:Using an ORTEC Gamma-X (GMX) detector, N-type crystal, the outer contact stage is only 0.3 microns boron thick.
    Standard P-type probe:P-type detectors, such as the GEM detector and the optimized GEM-M detector, have a thicker external contact stage (~600 micron lithium implant layer). This contact stage is three times thicker than that of the Gamma-X N-type detector. This results in a greatly increased probability that scattered rays will be blocked. 
    Energy extended P-type high-purity germanium detector:Energy-extended P-type detectors, such as the optimized GEM-C, only have an ultra-thin entrance window at the front of the detector. But the sides are surrounded by a thicker lithium ion implantation layer, so standard P-type detectors should not be used.
    High purity germanium crystal support cup Low Density Support Cup:ORTEC uses a 0.5mm thick, low-background aluminum support cup. In standard detector manufacturing.
    Copper material support cup:The use of copper as a crystal support cup should be avoided due to its high material density and mass absorption coefficient. This greatly reduces the probability of scattered photons entering the detector.
    Detector end cap Low Density End Caps: 1.5mm thick carbon fiber end caps or low background aluminum end caps with carbon fiber windows.
    Magnesium end cap:Magnesium material has a higher mass absorption coefficient than aluminum material, which greatly increases the probability of interaction between rays and matter. This reduces the probability of scattered rays entering the detector.
    Copper end cap: As with copper support cups, copper end caps should be avoided in low-background counting systems.
    Air between high purity germanium detector and shield, other materials
    Largest possible HPGe crystal diameter:In a typical 83mm diameter end cap, a 70mm diameter detector crystal can be placed. This is equivalent to placing an N-type detector with a relative efficiency of 70%. For custom Compton suppression systems, larger HPGe crystals and end caps can be used for improved performance.
     
    Fits detector size The HPGe detector end cap diameter should match the inner diameter of the annular NaI: When ordering, you need to order a NaI ring detector that matches the size of the detector end cap.
     

     
    Users familiar with low background detectors will note that magnesium end caps, copper end caps, and copper crystal support cups are commonly used with low background detectors. Although conventional thinking holds that Compton suppression systems are the solution for ultra-low activity sample measurements, this is not the case. By using these materials in the detector structure, the overall effect is to reduce the performance of the Compton suppression system (i.e., reduce the peak-to-complexity ratio). So in Compton suppression systems, trade-offs are very important.

    Other important components in the Compton suppression system
    In order to be able to simultaneously detect the Compton scattered rays scattered from the high-purity germanium detector into the NaI ring or plug detector and minimize the Compton platform height, the correct electronic setup is very important. Historically, ORTEC has provided a simulation system based on NIM plug-in electronics. With the rapid development of digital technology, ORTEC has launched a digital dual-channel MCA. The digital version of the Compton suppression system is significantly less complex and time-consuming to commission.

    The digital version of CSS significantly reduces the requirements for electronic equipment, using fewer components and a more compact space design. Large NIM racks are no longer required. In order to further improve space utilization without using any NIM electronics, ORTEC replaced the original box-type shielding body with a more integral integrated cast cylindrical shielding body. The shield is 4 inches (10 cm) thick and has an interior height that is 4 inches (10 cm) higher than traditional shields. The weight is only half of the original shielding body, and the floor space is also half of the original shielding body. In addition, the top sliding door design of the cylindrical shield allows the compliance detector to be installed in the top shield. This innovative approach greatly simplifies the loading and unloading of samples, requiring only one simple step, saving time. At the same time, it also reduces possible damage to the detector each time the sample is loaded and unloaded. Removing the plug detector from the NaI ring can greatly increase the sample measurement space without sacrificing the 4-π coverage geometry and Compton suppression performance.
    ORTEC Analog Compton Suppression SystemORTEC Digital Compton Suppression System

  • Performance and ordering information +


    The table below represents typical CSS options. 

    Upgradeable anti-cosmic ray functionality using plastic scintillator detectors. Please contact factory for details.

    For custom HPGe detectors (larger GMX detectors or planar detectors), annular coincidence detectors (larger inner diameter, different scintillation materials such as BGO), shields, or tables, please contact the factory.

    key components Model CSS-D-60
    detector The standard configuration is an N-type HPGe (GMX) 60% relative efficiency detector, using an 83mm diameter end cap, a low background Al support cup, and an integral carbon fiber end cap. Other sizes of detectors are available, with relative efficiency of 60% and end cap size of 83mm. 
    NaI ring and coincidence detector 9”x9” NaI ring, 3.5” inner diameter. 10.75” outer diameter. Resolution: @662 keV ≤11%.
    6”x2” NaI pancake detector. Resolution: @662 keV ≤7%.
    Sample size 3.5" diameter, 3" height (capacity up to 750 mL).
    electronics DSPEC-502A digital dual-channel MCA with coincidence/anti-coincidence function. Setup time is less than 30 minutes. The 556H high voltage module provides high voltage for the NaI ring detector.
    shield
    Cylindrical 4-inch (10 cm) thick low-background lead chamber with Tn/Cu composite shielding design on the inner layer. 
    Shield interior space: 11" diameter, 18" interior height.
    Optional anti-cosmic ray components (plastic scintillator detector).
    Software Contains GammaVision Gamma Spectroscopy Software. Normally not equipped with computer, if required please request when ordering.
    Performance guarantee value, peak-to-combustion ratio (ratio to the average height of the 500KeV accessory platform)
    1300:1 or better (using standard GMX60 detector). 
    Note: The P:C performance of using smaller GMX detectors will be relatively reduced.
    background count rate
    0.5 cps or less, use box or cylindrical factory shield.

    For a complete system, if specially designed shields, time conforming electronics, conforming detectors, high purity germanium detectors, etc. are required, please contact the factory.