EPD automatic irradiator EPD IRR-2

Survive in development, continuously optimize, and promote rapid development of enterprises with good reputation and scientific management

EPD automatic irradiator EPD IRR-2 image

EPD automatic irradiator EPD IRR-2

Product model:

Brief description:EPD automatic irradiator EPD IRR-2, with clever design and compact structure, can load 10 EPDs at a time without additional shielding devices. EPD IRR-2 has built-in β source and γ source and can detect EPD G and EPD MK2. The EPD loaded each time must be of the same model.

  • Manufacturer strength

    Manufacturer Strength
  • Valid warranty

    Valid Warranty
  • Quality assurance

    Quality Assurance

Products

PRODUCT CATEGORY

Related articles

RELATED ARTICLES

Detailed introduction

EPD automatic irradiator EPD IRR-2 Can automatically irradiate EPD MK2,EPD G。

Ÿ Built-in β Source andγ source

Ÿ Various detection methods

– with reference EPD compare

– Compare with previous measurements

– Compare doses to counts

Ÿ Compact structure and clever design

Ÿ No additional shielding required

Ÿ Detectable every time 10 a EPD

Ÿ Automate detection

Ÿ Detect simultaneously β、γ Measuring channel

EPD automatic irradiator EPD IRR-2,Ingenious design, compact structure, can be loaded every timeenter 10 a EPD, no additional shielding device is required.

EPD IRR-2 Built-in β Source and γ source, can be detected EPD G with EPD MK2。loaded each time EPD Must be of the same model.

The software has a friendly interface, simple operation and automatic execution. EPD Detection, supports multiple detectiontest mode. The software automatically saves measurement data to facilitate users to perform queries or other operations.

Instrument performance

Irradiation source: γ: Am-241, 3.7GBq(100mCi)

β: Cl-36, 100kBq(2.7 μCi)

Performance 241Am 60keV, the deviation is less than±2%, measurement time 2 minutes

(15℃~25℃, confidence level 95%)

36Cl 714keV

(The statistical error is less than 10%, measuring time 2 minutes)

leak Under normal operation, distance from any surface 50mm The dose rate at is less than 1μSv/h

environment

working temperature 5℃~40℃

Storage temperature -25℃~70℃

instructions

indicator light Power on status (yellow);The radioactive source has been shielded (green); Irradiating (red); machinery

Size 470×400×320mm

weight 37kg

power supply

Voltage 240V±10%, 47~63Hz

Power consumption 300W

Power outage handling power outage 5s Automatically terminate measurement after

fuse 2A T,20×5mm(diameter)

connect IEC 320

EMC BS EN 61010-1

Product consultation