KARRAY / PRODUCT DETAILS

RD-120 Backpack Isotope Identifier

Survive in development, continuously optimize, and promote rapid development of enterprises with good reputation and scientific management

RD-120 Backpack Isotope Identifier image

RD-120 Backpack Isotope Identifier

Product model:

Brief description:RD-120 series of backpacks can be customized for specific applications with a variety of detectors and network options Low power drain allows for multiple hours of uninterrupted monitoring Allows other user software that may be deployed to "command and control" type CONOPS

  • Manufacturer strength

    Manufacturer Strength
  • Valid warranty

    Valid Warranty
  • Quality assurance

    Quality Assurance

Products

PRODUCT CATEGORY

Related articles

RELATED ARTICLES

Detailed introduction

RD-120 Backpack Isotope Identifier

Product introduction:

l RD-120 series of backpacks that can be customized for specific applications with a variety of detector and network options

l Low power consumption allows multiple hours of uninterrupted monitoring

l Allow other users software may be deployed to"Command and Control" CONOPS

l Alarm options include smartphone app and standing position

l The wearer can use the backpack to zero in on possible causes of alarm, take one or more pictures of an area with a smartphone, and send those pictures back to the command center to supplement the data automatically transmitted to the backpack

Product features:

l Gamma and neutron detection/ ID

l Backpack, Zhiying or door configuration

l Real-time isotope identification

l Bluetooth notification

l text/voice communication

l Command center support

l IP-67 waterproof (wan fully purified)

Product application:

l emergency response

l law enforcement

l border security

l Industry

l medical

l radiation safety

l Portal monitoring

l Passenger and freight monitoring

l non-proliferation enforcement

l health physics

l Environmental waste monitoring

Technical parameters:

detector

NAI , 6Lil or CEBR

integrated electronics

digital signal processingMCA

energy range

18 keV - 3 MeV

Controller display

320×240 high brightness 32000 colors 3.5" translucent LCD display

controllerI / O

10/100 Ethernet port and CompactFlash card reader with USB adapter

power supply

8 standard AA batteries

weight

4.5 lbs. 2" x 2" NaI detector and battery

Size

12"L x 4"(H) x 10"W (not including detector)

water/dustproof level

IP56

temperature range

-20~ 50 °C

control

Use your thumbs to operate the keyboard7 key customization

Alert

Visual (on screen) and audio (built-in speakers or optional headphones)

Expertise

Secondary compression conversion ( QCC) allows mixed isotope identification

lag

provide97% I.D. within 2 seconds confidence level

Optional modules

serialGPS Receiver, Wireless Communications Spectrum Report Mapping

14-bit ADC

Type: Basic Conversion Pipeline Flash

mode

Linear256, 512, 1024, QCC 256, 512 (US patent 5608222)

Calibration

Automatic temperature stabilization

Customized

Isotopes and their associated energy lines, whether modified or used in the field

Product consultation

Message box

Please enter the calculation result, such as: three plus four = 7

After submission, we will receive your requirements. You can also download the requirement file or call 021-20989890 directly.